The training and research facility based on three desktop powder diffractometers Rigaku «MiniFlex II»
Desktop powder diffractometers for qualitative and quantitative phase analysis of polycrystals within the θ / 2θ scanning geometry using Cupper or Cobalt characteristic wavelengths
Technical charachteristics:
- CuKα and CoKα X-ray sealed tubes (focus size - 1.0 x 10 mm);
- Rated operation X-ray sources: 30 kV/10 mA;
-
Vertical Theta / 2Theta goniometer radius of 150 mm:
- 2θ angle ranges from -3 to +145 °;
- Scanning method – related θs / θd;
- Maximum scanning speed 100 ° / min;
-
High Energy-detector of the reflected X-rays DTEX / ULTRA:
- 512 recording channels;
- dynamic range of at least 105 pulses / s per channel;
-
Optical System:
- manual gap on the output beam;
- manual gap on the diffracted beam;
- manual receiving slit;
Software:
The PDXL software package - provides an opportunity for a different analysis, for example, automatic qualitative phase analysis, quantitative phase analysis, refining unit cell parameters.
Information from the manufacturer:
http://www.rigaku.com/products/xrd/miniflex
Location:
The Vasilyevsky Island training and research Complex of SPbSUBack to equipment list