The research facility with low- and high-temperature cameras based on Rigaku «Ultima IV»
A high-resolution diffractometer with low- and high-temperature cameras additionally equipped with a differential scanning calorimetry system for studying phase transformations within the temperature range of 93 K to 1773 K.
Technical charachteristics:
- CuKα X-ray sealed tube (focus size - 0.4 x 12 mm);
- Rated operation X-ray sources: 40 kV/35 mA;
- Vertical Theta / Theta goniometer radius of 185 mm:
- 2θ angle ranges from -3 to +162 °;
- Scanning method – related θs / θd or θs, θd independent;
- min scanning step value 0.0001°;
- High Energy-detector of the reflected X-rays DTEX / ULTRA:
- 512 recording channels;
- dynamic range of at least 105 pulses / s per channel;
- Optical System:
- manual gap on the output beam;
- manual gap on the diffracted beam;
- manual receiving slit;
- automatic adjustment of the tube height, goniometer, optics and detector;
- Medium and low-temperature chamber Rigaku “R-300”:
- resistant, external to the sample, type of heater;
- sample size of at least 17mm x 25mm x 2.0mm;
- independent thermocouple and heater design;
- temperature gradient across the sample does not exceed ± 2.5 ° C for temperatures not exceeding 100 ° C;
- operating temperature in a vacuum ranges from -180 ° C to 300 ° C;
- material of the sample holder – Cu;
- High-temperature chamber Rigaku “SHT-1500”:
- resistant, external to the sample, type of heater;
- sample size of at least 11mm x 25mm x 0.5mm;
- independent thermocouple and heater design;
- temperature gradient across the sample does not exceed ± 0.5 ° C for up to 500 ° C temperatures and ± 5 ° C for above 1400 ° C temperatures;
- max operating temperature in vacuum of at least 1450 ° C;
- max operating temperature in the air at least 1500 ° C;
- max operating temperature in an inert gas (He) is not less than 1300 ° C;
- material of the sample holder – Pt;
Software:
The PDXL software package - provides an opportunity for a different analysis, for example, automatic qualitative phase analysis, quantitative phase analysis, refining unit cell parameters.
Information from the manufacturer:
http://www.rigaku.com/products/xrd/ultima
Location:
The Vasilyevsky Island training and research Complex of SPbSUBack to equipment list